Product Introduction
Hexa Pharma Chem presents the Integtest® Glove Integrity Tester GT-3.0, manufactured by Beijing Neuronbc Laboratories Co., Ltd. The GT-3.0 is an online glove integrity testing instrument designed according to the leak detector principle recommended by GB/T 25915.7-2010 / ISO 14644-7:2004. It is a dedicated instrument for detecting leakage of gloves in isolators, RABS, and cleanroom barrier systems.
The GT-3.0 features a simple and easy design that requires no additional equipment and no need to remove gloves, delivering fast detection speed with high precision. It includes a color touch screen with humanized interface design, automatic program selection based on test parameters, authority management with electronic signature compliant with 21 CFR Part 11, built-in printer for historical records, and USB data export capability. An ideal solution for pharmaceutical aseptic processing and barrier integrity validation.
Product Features
- Small-volume frequent air inlet — uniform velocity avoids glove deformation
- Color touch screen with humanized interface design
- Automatic selection of best-fit test program based on glove parameters
- Date and time display for complete test traceability
- Authority management & electronic signature — 21 CFR Part 11 compliant
- Convenient to move, easy to carry, quick connection
- Historical records — print, storage, and query functionality
- No additional equipment required — simple and easy design
- No need to remove gloves during testing
- Fast detection speed with high precision results
- Built-in printer for immediate test result documentation
- USB data export for audit trail (customization available)
- Replaceable glove trays — supports different glove specifications
- Li-ion battery with oil-free, water-free pump option
Application Areas
Technical Parameters
| Working Power Supply | AC 220V, 50Hz, 120W (with battery) |
| Rated Power / Test Duration | 2–8 minutes |
| Pressure Range | 500–5000 Pa |
| Response Value | ≤ 50 Pa |
| Noise | ≤ 50 dB(A) |
| Working Mode | Continuous Work |
| External Gas Source | Compressed Air, Nitrogen, Inert Gas |
| Sealing Pressure | ≤ 0.8 Mpa |
| Other Functions | Audit Trail, Data Export by USB (customization) |
| Historical Record | 10,000 groups |
| Print Function | Built-in Printer |
| Dimensions | 400 × 260 × 270 × 250 mm |
| Weight | 10 KG |
| Online Gloves Tray | Free supply 1 size; customization available |
| Battery | Li-ion Battery |
| Pump | Oil-free, Water-free (built-in pump customizable) |
Product Content — Application Details
- Sterile Inspection of Purified Water & WFI
- Microbial Limit Testing for Sterile Preparations
- Isolator Glove Integrity Validation
- RABS Barrier System Testing
- Large Infusion & Small Injection Testing
- Biological & Blood Product Applications
- Syringe, Infusion Set & Transfusion Set Testing
- IV Catheter Integrity Verification
For Our Products & Services Contact Us